Show simple item record

dc.contributor.authorWatson, Robin.J.
dc.contributor.authorBaranwal, Vikas.C.
dc.contributor.authorOfstad, Frode
dc.contributor.authorDumais, Marie-Andrée
dc.date.accessioned2024-06-28T07:49:47Z
dc.date.available2024-06-28T07:49:47Z
dc.date.issued2024
dc.identifier.issn0800-3416
dc.identifier.urihttps://hdl.handle.net/11250/3136449
dc.description.abstractNGU owns and uses a set of transportable concrete calibration pads for calibration of its airborne gamma-ray spectrometers. The pads are principally designed for hand-held instruments, in contrast with larger fixed pads which are specially designed for airborne systems. Such a set of fixed pads are maintained by the Swedish Geological Survey (SGU) at Dala airport, Borlänge. In August 2012 NGU's airborne system was calibrated using the Borlänge pads, allowing a comparison between transportable and fixed calibration pads. A comparison of the two calibration methods shows little significant differences. As such NGU's calibration pads are a suitable and convenient substitute for dedicated fixed calibration pads. Calibration parameters obtained in recent years from the NGU pads are also provided as a monitor of long-term instrument stability.
dc.language.isoeng
dc.publisherNorges geologiske undersøkelse
dc.relation.ispartofseriesNGU-Rapport (2024.016)
dc.rightsNavngivelse 4.0 Internasjonal
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/deed.no
dc.subjectRADIOMETRI
dc.subjectSPEKTROMETRI
dc.subjectHELIKOPTERMÅLING
dc.subjectFLYMÅLING RADIOAKTIVITET
dc.subjectGAMMASPEKTROMETRI
dc.titleA comparison of transportable and fixed calibration pads for airborne gamma-ray spectrometers
dc.typeReport
dc.description.localcode69623
dc.source.pagenumber12 s.
dc.relation.project(335900) Atomberedskap